Basic accuracy: Experience precision at its finest with an outstanding basic accuracy of ±0.04%, reaching up to a remarkable maximum of ±0.025%.
Range ratio: Impressive flexibility with a comprehensive range ratio of 100:1, catering to diverse application needs.
Response time: Experience lightning-fast data processing with a response time of just 90ms and a data refresh time of 50ms, ensuring real-time accuracy.
Range range: Offers a robust range for differential pressure, spanning 0-100 Pa to 14MPa, perfect for varied industrial requirements.
Pressure of 0-1 kPa-70MPa: Seamlessly accommodates pressure levels from 0-1 kPa to 70MPa, ensuring reliable performance across a broad spectrum.
Output signal: Enjoy seamless communication with a 4-20 mA output signal that features superimposed HART protocol communication, compatible with both HART6.0 and HART 7.0 versions.
PROFIBUS-PA fieldbus communication: Connect effortlessly with the PROFIBUS-PA fieldbus communication system, utilizing version 3.02 for enhanced data transfer.
FF Fidbus Communication: Benefit from efficient networking with FF Fidbus Communication, operating with the reliability of Version 6.1.1.
Overload limit: Capable of enduring an overload limit of 42MPa for differential pressure, ensuring robustness in demanding applications.
Pressure of 105MPa: Handles pressure of up to 105MPa, offering exceptional durability and reliability.
Effect of the static pressure: Minimal impact with a static pressure effect of just 0.03%FS per 16MPa, maintaining precision under varied conditions.
Temperature range: Flexible ambient temperature range from -40ºC to 80ºC, with specific considerations for explosion-proof applications.
Process temperature: Operates efficiently across a wide process temperature range from -90ºC to 700ºC, ensuring adaptability to extreme conditions.
LCD temperature: Keeps its cool within a temperature range of -35ºC to 80ºC, ensuring clear displays under varying conditions.
Long-term drift: Maintains exceptional stability with a long-term drift of ±0.1%FS over 10 years, promising reliability and peace of mind.
Temperature Effect: Temperature effect is finely tuned with a formula of (0.06 * r + 0.01)% per 28ºC, covering a range from -40ºC to 80ºC, ensuring consistent performance.
Explosion-proof grade: Certified with NEPSI for Type Ex iaIICT4 Ga, suitable for ambient temperatures from -40ºC to 60ºC, ensuring safety in hazardous environments.
NEPSI: Offers intrinsic safety with Ex iaIICT6 Ga certification, fitting ambient temperatures from -40ºC to 40ºC, catering to specific safety needs.
NEPSI: Designed for explosive conditions with type Ex dIICT6 Gb certification, applicable from -40ºC to 60ºC, ensuring safety in volatile settings.
NEPSI: Dust explosion safety guaranteed with Ex tD A21 IP67 T85ºC certification, offering peace of mind in dusty environments.
NEPSI: 'n' explosion-proof design with Ex nACT6 Gc certification, suitable for temperatures from -40ºC to +60ºC, providing versatile safety solutions.
ATEX: Certified with Ben Ex II 1 G Ex iaIICT4 Ga, ensuring compliance with stringent safety standards.
ATEX: Burst protection with Ex II 1 / 2 G Ex dbIICT6-T 4 Ga / Gb certification, offering robust safety features for explosive atmospheres.
IECEx: Guaranteed intrinsic safety with Ex iaIICT4 Ga certification, providing international safety assurance.
IECEx: Burst protection certified with Ex dbIICT6-T 4 Ga / Gb certification, ensuring safe operation in explosive environments.